Image-based kidney dosimetry for [177Lu]Lu-DOTATATE treatments Development, validation, and assessment of quantification accuracy and precision
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
This thesis aims to develop, validate, and evaluate methods for accurate and precise image-based kidney dosimetry in patients with neuroendocrine tumours (NETs) treated with [177Lu]Lu-DOTATATE. Special focus is placed on quantifying and reducing uncertainties in kidney dose estimation. In paper I, an automatic segmentation method for the whole kidney parenchyma (WKP) was developed using a convolutional neural network (CNN). The CNN-based segmentation demonstrated performance comparable to manual (CT-based) segmentation, with a strong correlation in activity concentration measurements (r > 0.96, p < 0.01) indicating its suitability as an alternative for clinical dosimetry workflows. In paper II, the small-volume-of-interest (SV) method was optimized by adjusting VOI size and increasing the number of VOIs per kidney. This reduced dose estimation uncertainty from 14% when using a single 4 mL SV to 8.3% when using five 2 mL SVs, compared to the WKP method used as the reference. Papers II & III established a clinically feasible routine for calculating patient-specific recovery coefficients (RCs) using Monte Carlo simulations, reducing RC-related uncertainty by up to 4% compared to the fixed RC of 0.85 recommended by EANM guidelines. Paper III quantified anatomical changes in kidney volume (up to 10.77%) during treatment due to amino acid infusion. These changes were shown to result in absorbed dose overestimation of up to 2.5% when a single-timepoint CT segmentation was propagated across serial SPECT scans. In paper IV, a novel framework was introduced that uses an empirical power-law model to quantify the accuracy of WKP-based kidney dosimetry using SV data. The model demonstrated a relative precision of 7.27% for 2 mL SV and 6.92% for 0.6 mL SVs.
In conclusion, this work presents a systematic evaluation of image-based kidney dosimetry methods, identifying key sources of uncertainty and demonstrating how they can be mitigated through optimized VOI strategies, patient-specific recovery correction, and CNN-based segmentation. A quantitative framework for estimating relative accuracy was also introduced, supporting the development of robust and clinically feasible dosimetry practices for individualized molecular radiotherapy in NET patients.
Description
Keywords
Dosimetry, 177Lu-DOTATATE, Molecular radiation therapy