Cone Beam Computed Tomography - radiation dose and image quality assessments

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

Diagnostic radiology has undergone profound changes in the last 30 years. New technologies are available to the dental field, cone beam computed tomography (CBCT) as one of the most important. CBCT is a catch-all term for a technology comprising a variety of machines differing in many respects: patient positioning, volume size (FOV), radiation quality, image capturing and reconstruction, image resolution and radiation dose. When new technology is introduced one must make sure that diagnostic accuracy is better or at least as good as the one it can be expected to replace. The CBCT brand tested was two versions of Accuitomo (Morita, Japan): 3D Accuitomo with an image intensifier as detector, FOV 3 cm x 4 cm and 3D Accuitomo FPD with a flat panel detector, FOVs 4 cm x 4 cm and 6 cm x 6 cm. The 3D Accuitomo was compared with intra-oral radiography for endodontic diagnosis in 35 patients with 46 teeth analyzed, of which 41 were endodontically treated. Three observers assessed the images by consensus. The result showed that CBCT imaging was superior with a higher number of teeth diagnosed with periapical lesions (42 vs 32 teeth). When evaluating 3D Accuitomo examinations in the posterior mandible in 30 patients, visibility of marginal bone crest and mandibular canal, important anatomic structures for implant planning, was high with good observer agreement among seven observers. Radiographic techniques have to be evaluated concerning radiation dose, which requires well-defined and easy-to-use methods. Two methods: CT dose index (CTDI), prevailing method for CT units, and dose-area product (DAP) were evaluated for calculating effective dose (E) for both units. An asymmetric dose distribution was revealed when a clinical situation was simulated. Hence, the CTDI method was not applicable for these units with small FOVs. Based on DAP values from 90 patient examinations effective dose was estimated for three diagnostic tasks: implant planning in posterior mandible and examinations of impacted lower third molars and retained upper cuspids. It varied between 11-77 μSv. Radiation dose should be evaluated together with image quality. Images of a skull phantom were obtained with both units varying tube voltage, tube current, degree of rotation and FOVs. Seven observers assessed subjective image quality using a six-point rating scale for two diagnostic tasks: periapical diagnosis and implant planning in the posterior part of the jaws. Intra-observer agreement was good and inter-observer agreement moderate. Periapical diagnosis was found to, regardless of jaw, require higher exposure parameters compared to implant planning. Implant planning in the lower jaw required higher exposure parameters compared to upper jaw. Substantial dose reduction could be made without loss of diagnostic information by using a rotation of 180°, in particular implant planning in upper jaw. CBCT with small FOVs was found to be well-suited for periapical diagnosis and implant planning. The CTDI method is not applicable estimating effective dose for these units. Based on DAP values effective dose varied between 11-77 μSv (ICRP 60, 1991) in a retrospectively selected patient material. Adaptation of exposure parameters to diagnostic task can give substantial dose reduction.

Description

Keywords

Cone beam computed tomography, anatomic landmarks, dose-area product, image quality, implant planning, periapical diagnosis, radiation dosimetry

Citation

ISBN

978-91-628-8205-1

Articles

I Lofthag-Hansen S, Huumonen S, Gröndahl K, Gröndahl HG Limited cone-beam CT and intraoral radiography for the diagnosis of periapical pathology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007 Jan;103(1):114-9. Epub 2006 Apr 24. ::PMID::17178504

II Lofthag-Hansen S, Gröndahl K, Ekestubbe A. Cone-beam CT for preoperative implant planning in the posterior mandible: visibility of anatomic landmarks. Clin Implant Dent Relat Res. 2009 Sep;11(3):246-55. Epub 2008 Sep 9. ::PMID::18783419

III Lofthag-Hansen S, Thilander-Klang A, Ekestubbe A, Helmrot E, Gröndahl K. Calculating effective dose on a cone beam computed tomography device: 3D Accuitomo and 3D Accuitomo FPD. Dentomaxillofac Radiol. 2008 Feb;37(2):72-9. ::PMID::18239034

IV Lofthag-Hansen S, Thilander-Klang A, Gröndahl K. Evaluation of subjective image quality in relation to diagnostic task for cone beam computed tomography with different fields of view. Eur J Radiol. 2010 Oct 19. [Epub ahead of print] ::PMID::20965675